Un couple de frottement métal-métal peut-il améliorer la longévité des prothèses totales de hanche?

Dimension: px
Commencer à balayer dès la page:

Download "Un couple de frottement métal-métal peut-il améliorer la longévité des prothèses totales de hanche?"

Transcription

1 Revue de chirurgie orthopédique Masson, Paris, , 91, MISE AU POINT Un couple de frottement métal-métal peut-il améliorer la longévité des prothèses totales de hanche? Can metal-on-metal bearings improve the longevity of total hip prostheses? C. Delaunay Clinique de l Yvette, 43, route de Corbeil, Longjumeau. ABSTRACT Purpose of the study In total hip arthroplasty (THA), inter-series comparative clinical results cannot be considered worthy before at least 10 years of average follow-up, as shown in the Swedish Arthroplasty Register experience ( ). Last generation metal-on-metal bearings were introduced in France only nine years ago (1995). To date, using mid term information, data from the literature, and our experience, one could assume that this bearing material has the capacity to improve THA longevity. Three types of information were analyzed: 1) comparative radiographic and EBRA studies published on early migration (< 2 years) of acetabular implants; 2) preliminary comparative data on wear and osteolysis at the 8-year maximum follow-up; 3) clinical data on dislocation frequency and in vitro and in vivo observations of bearing separation and sliding. Study n o 1: primary stability of these metal-on-metal acetabular cups was better than for polyethylene (PE) cups or alumina liners; study n o 2: no detectable wear and less osteolysis were observed to date with metal-on-metal bearings versus their alumina-on-pe counterparts. Study n o 3: less dislocation and less head sliding were measured with the metal-on-metal versus alumina-on-pr bearings possibly due to the interfacial forces provided by lubricating fluid (suction fit). Conclusion According to current knowledge, this metal-on-metal bearing still represent, with alumina-on-alumina and highly crosslinked PE, a competitive and comprehensive option to improve THA longevity. The real pending problem concerns the frequency and above all the intensity of metal delayed hypersensitivity reaction and their potential effect on implant loosening. Key words: Total hip arthroplasty, bearings, metal-on-metal. RÉSUMÉ En arthroplastie totale de hanche, aucune comparaison entre les résultats cliniques de divers implants n est raisonnable avant d avoir franchi le seuil de 10 ans de recul moyen, comme cela a bien été montré dans l expérience du registre suédois ( ). Le couple de frottement métal-métal (Metasul) n ayant été introduit dans sa forme actuelle en France que depuis 9 ans (1995), sur quels arguments intermédiaires, provenant de la littérature ou de l expérience de l auteur, peut-on à ce jour apprécier la capacité réelle de ce matériel à améliorer la longévité des arthroplasties qui en sont équipées? Trois groupes d informations ont été analysés : 1/ les travaux comparatifs radiographiques et EBRA sur les migrations précoces des implants acétabulaires, qui indiquent que la stabilité initiale primaire de ces cupules métal-métal est meilleure que celle des cupules avec inserts en polyéthylène ou alumine ; 2/ les données préliminaires au recul maximum disponible de 8 ans, qui ne montrent toujours aucune usure et moins d ostéolyse avec ces PTH à couple métallique versus alumine-pe ; 3/ les études des luxations et les observations obtenues in vitro et in vivo sur les phénomènes de piston-décoaptation, qui indiquent une réduction de la fréquence des luxations et de l intensité de la décoaptation avec les couples métal-métal versus alumine-pe, possiblement liée à un phénomène d adhésivité «par succion» du au film lubrifiant. Tirés à part : C. DELAUNAY, à l adresse ci-dessus. c.delaunay@clinique-yvette.com Acceptation définitive le : 30 août 2004

2 UN COUPLE DE FROTTEMENT MÉTAL-MÉTAL PEUT-IL AMÉLIORER LA LONGÉVITÉ DES PROTHÈSES TOTALES DE HANCHE? 71 Dans l état actuel des connaissances, le couple métal-métal représente toujours, avec les couples alumine-alumine et les PE fortement réticulés, une option alternative susceptible d améliorer la longévité des PTH. Le réel problème en suspens est représenté par la fréquence et surtout l intensité des réactions immunoallergiques d hyper-sensibilité retardée aux débris métalliques et leur implication potentielle dans la genèse des descellements prothétiques. Mots clés : Prothèse totale de hanche, couple de frottement, métal-métal. INTRODUCTION En arthroplastie totale de la hanche, rien de sérieux en ce qui concerne les résultats cliniques ne peut être avancé avant un recul minimum de 10 ans. Ainsi, depuis sa création en 1978, il a fallu attendre 15 années avec la première publication de 1993, pour que le registre suédois des prothèses totales de la hanche (PTH) puisse mettre enfin en évidence des différences significatives entre la survie des différents implants étudiés [Malchau et al. (1)]. Depuis, les rapports successifs du registre suédois ont toujours identifié le descellement aseptique comme la première cause d échec des PTH. La dernière version de 2002 indique par ailleurs que la probabilité de révision pour descellement aseptique des PTH à couples «traditionnels» (utilisant du polyéthylène acétabulaire conventionnel en frottement), implantées pour coxarthrose primitive et cimentées selon les techniques modernes de seconde génération, est encore de 4,8 % à 10 ans et atteint même 10,1 % à 10 ans chez les patients âgés de moins de 55 ans [Malchau et al. (2)]. L ostéolyse péri-prothétique est un processus réactionnel complexe, induit par tous les débris et dépendant de leur quantité et surtout de leur taille. Elle est universellement considérée comme la cause essentielle du descellement et le polyéthylène (PE) conventionnel producteur de débris ostéolytiques, comme le «maillon faible» de nos arthroplasties. Cette constatation a conduit à la recherche logique de couples prothétiques supprimant le PE en frottement, d abord en France avec la combinaison céramique d alumine-sur-céramique d alumine (Boutin, 1971), puis en Suisse avec la réintroduction d un couple CoCr-sur-CoCr de seconde génération, le Metasul (Zimmer-Centerpulse, Winterthur, CH) par Weber en La fabrication de ces couples «dur-sur-dur» nécessite une compétence technologique toute particulière, le «jeu» articulaire prothétique (clearance) jouant un rôle fondamental dans la qualité du frottement, condition sine qua non à la réduction de la production des débris d usure phagocytables (taille de 0,3 à 10 microns) et dès lors potentiellement ostéolytiques. En , l auteur avait fait un état des lieux sur les couples de frottement [Delaunay (3, 4)]. Depuis cette époque, le couple métal-métal représente toujours, avec le couple alumine-alumine et plus récemment avec la mise au point de PE hautement réticulés, une des trois options disponibles susceptibles de réduire la production de débris d usure ostéolytiques et d améliorer la longévité des implants, ce qui vient d être récemment confirmé Outre- Atlantique [Heisel et al. (5)]. Mais il faudra, pour chacune d entre elles, attendre un recul moyen de plus de 10 ans avant d avoir la possibilité d en apprécier cliniquement la validité par la mise en évidence d une amélioration significative de la survie des arthroplasties qui en sont équipées par rapport aux données connues des PTH à couples de frottement «conventionnels». Or, le couple Metasul n a été introduit sur le marché français sous sa forme actuelle qu en 1995 (année de son marquage CE), en remplacement de la première version avec billes à jupe commercialisée en Suisse en La problématique ici posée est donc simple : après huit années seulement d expérience en France, sur quels arguments cliniques peut-on aujourd hui s appuyer afin de soutenir l hypothèse selon laquelle un couple métal-métal tel que le Metasul dans sa version actuelle, présente une potentialité réelle à améliorer la longévité des arthroplasties totales de hanche qui en sont équipées? MATÉRIEL ET RÉSULTATS Les éléments de réponse tiennent en trois points : l étude de la migration précoce des implants acétabulaires ; la détection radiographique de l usure des implants et de la survenue des manifestations ostéolytiques ; et la détermination de la fréquence des accidents de luxation corrélée à l étude cinétique in vivo des phénomènes de piston-décoaptation survenant entre les composants en frottement. La migration précoce des implants acétabulaires En 1995, les travaux de Freeman [Stocks et al. (6)] aboutissaient à la constatation que la migration radiologique détectable des implants acétabulaires (non cimentés) survenant dans les deux années suivant l implantation était un signe péjoratif annonciateur d un «descellement» secondaire et d une fréquence accrue des révisions de ces implants précocement dépistés «instables». Ces auteurs avaient ainsi établi une relation quasi-linéaire entre intensité de la migration précocement détectable et la fréquence des révisions des implants acétabulaires non cimentés étudiés (fig. 1). Ainsi, pour une migration proximale moyenne de 2 millimètres, mesurée dans les deux premières années après implantation, le taux de révision des cupules concernées par cette migration était de l ordre de 30 % au recul de 6 ans. La méthode EBRA (Ein Bild Röntgen Analyse), essentiellement développée dès la fin des années 1990 par T. Ilchmann en Autriche, s est avérée beaucoup plus perfor-

3 72 C. DELAUNAY FIG.1. Diagramme de la correspondance entre la migration proximale moyenne des cupules à 2 ans (ordonnée, en mm) et les taux de révision correspondants au-delà de la 6 e année postopératoire (abscisse, en %) [extrait de l article de Stocks et al. (6), avec permission]. mante que les radiographies standards pour détecter les migrations précoces des cupules des PTH, de même que pour mesurer l usure des implants en PE [Ilchmann et al. (7)]. Elle fut utilisée dans l étude autrichienne de Boehler et al. (8) comparant deux groupes de PTH identiques comportant des cupules modulaires non cimentées à l exception de leur couple de frottement, bille alumine-insert PE versus bille alumine-insert alumine. Dans les 2 premières années post-implantation, la méthode EBRA indiquait 4 fois moins de migration avec la combinaison alumine-pe et cliniquement au-delà de la 6 e année, une probabilité de révision acétabulaire pour descellement aseptique de 1,8 % ( 1,7) pour la combinaison alumine-pe, alors qu elle atteignait déjà 11 % ( 6) pour la combinaison alumine-alumine précocement moins stable, cette différence étant significative (p = 0,032). La valeur prédictive péjorative de la détection d une migration des implants acétabulaires dans les deux années suivant leur implantation suggérée par les travaux de Freeman, se voyait donc bien confirmée dans l étude EBRA de Boehler et al. (8). Tout récemment, une étude prospective et randomisée similaire a été menée par l équipe de Krismer à l université d Innsbruck [Pabinger et al. (9)]. Elle comparait là encore deux groupes de PTH identiques (tiges et cupules acétabulaires cimentées de Weber) à l exception de la nature de leur couple de frottement en 28 millimètres, CoCr-CoCr- Metasul versus céramique d alumine-pe. Dans les deux premières années après l implantation, la méthode EBRA mettait en évidence une nette tendance à une migration moindre des cupules avec la combinaison Metasul (0,4 mm versus 1,1 mm pour les cupules avec couple de frottement alumine-pe), mais cette différence n atteignait pas le niveau de significativité requis (p = 0,07). Les auteurs en concluaient cependant que le couple de frottement Metasul n entraînait pas de migration plus importante des cupules malgré un coefficient de friction théoriquement supérieur (ce qui était leur hypothèse de départ). La synthèse de ces données comparatives permet de constater que par la méthode performante EBRA et à calibre identique, la stabilité primaire des cupules acétabulaires cimentées des PTH équipées d un couple de frottement Metasul est supérieure à celle des cupules acétabulaires avec couple céramique d alumine-pe, elle-même supérieure à celle des cupules acétabulaires non cimentées à couple de frottement alumine-alumine. Cette meilleure stabilité initiale des cupules acétabulaires Metasul devrait selon la théorie de Freeman se traduire par une réduction de la fréquence des «descellements» et des révisions à plus long terme. Usure et ostéolyse Dans l étude précédente de Pabinger et al. (9), la méthode EBRA avait permis de mesurer une usure moyenne in vivo des couples alumine-pe qui atteignait déjà 0,72 mm à la fin de la 2 e année après implantation alors qu elle n était toujours pas décelable pour la combinaison Metasul, et ce malgré la finesse de détection de la méthode EBRA utilisée. Dans la population générale, nos études prospectives avec une PTH non cimentée à couple de frottement alumine-pe en 28 mm (Alloclassic, Zimmer-Centerpulse, Winterthur, CH) ont toujours démontré une excellente ostéointégration des composants. En revanche, la fréquence de survenue d une ostéolyse, quoique faible, ne cesse d augmenter avec le temps d observation, de 1 % à 6 ans de recul moyen à 3,4 % à 7,3 ans [Delaunay et al. (10, 11)]. De plus, il est bien acquis que l usure des couples de frottement des PTH est liée à l intensité de l usage qu en font les patients, logiquement corrélée à leur espérance de vie et à leur niveau d activité [Delaunay (4), Heisel et al. (5)]. Dans une étude multicentrique d une série de 115 PTH primaires non cimentées à couple de frottement alumine-pe en calibre 28 mm chez des patients de moins de 65 ans, Bonnomet et al. (12) retrouvaient après un recul moyen de 8,2 ans (de 5 à 12 ans) une usure globale de l insert en PE supérieure à 1 mm dans 9,6 % des cas et une probabilité de survenue d une ostéolyse fémorale confirmée de 3,9 % à 10 ans. Une de ces PTH avec ostéolyse étendue détectée dès la 5 e année a depuis été reprise, 11 ans après l intervention initiale (fig. 2). Dans une série personnelle prospective de 109 PTH consécutives du même modèle que celui utilisé dans l étude de Bonnomet et al. (12) mais avec un couple de frottement Metasul, implantées de mai 1995 à octobre 2002 chez nos patients de moins de 65 ans, nous retrouvons actuellement pour un recul moyen de 3,8 ans seulement (1 à 8 ans) : aucune usure radiologiquement encore décelable

4 UN COUPLE DE FROTTEMENT MÉTAL-MÉTAL PEUT-IL AMÉLIORER LA LONGÉVITÉ DES PROTHÈSES TOTALES DE HANCHE? 73 [ce qui vient confirmer à plus long terme les constations déjà faites à 2 ans avec la méthode EBRA par Pabinger et al. (8)] ; une hanche avec ostéolyse fémorale proximale en zone 1 de Gruen cliniquement asymptomatique ; et une probabilité de survie des tiges et des anneaux vissés pour l événement «descellement aseptique» de 100 % à 8 ans avec seulement 23 hanches «à risque» (données non publiées par manque de recul). À ce jour cependant, la comparaison entre ces deux études menées chez des patients de moins de 65 ans avec des PTH non cimentées voisines, indique qu au recul de 8 ans, les PTH à couple de frottement Metasul montrent peu de remaniements ostéolytiques fémoraux et encore aucune usure détectable, ce qui n est déjà plus le cas avec le couple alumine-pe 28 millimètres. La littérature est encore pauvre en résultats de séries cliniques avec le couple Metasul, qui n a rappelons-le que 8 ans d ancienneté dans sa forme actuelle. Signalons cependant les bons résultats de la séries de Yun et al. (13) avec 100 PTH Metasul, aucune ostéolyse inquiétante et une survie de 98,8 % à 11 ans pour l événement «révision pour descellement aseptique». Chez des patients de moins de 50 ans, les résultats de la série de Migaud et al. (14) sont prometteurs, encore sans aucun signe de descellement mais pour un recul moyen plus modeste de 5,1 ans seulement. Fréquence des luxations et phénomènes de «piston-décoaptation» FIG. 2. Ostéolyse fémorale et usure de l insert acétabulaire en polyéthylène avec excentration de la bille prothétique chez un patient âgé de 50 ans opéré 8 ans auparavant [extrait de l article de Bonnomet et al. (12)]. La survenue d une luxation représente un événement majeur dans l évolution des PTH dont la morbidité et les conséquences fonctionnelles restent pour les patients un sujet permanent de préoccupation. La réduction de leur fréquence repose sur de nombreux facteurs, techniques et démographiques, au sein desquels les facteurs tribologiques (en dehors du calibre des implants en frottement) sont encore mal appréciés. C est de façon relativement empirique que Clarke et al. (15) avaient constaté une apparente réduction de la fréquence des luxations de leur PTH après introduction dans leur pratique en 1998 de l usage d une cupule à couple métal-métal en 28 millimètres (Ultima, Johnson & Johnson), en remplacement du couple céramique-pe (à débord postérieur) de même calibre utilisé précédemment, mais toujours sur la même PTH. Ils étudièrent alors toutes les arthroplasties primaires réalisées de 1995 à 2002, en excluant certaines conditions cliniques connues pour favoriser l instabilité prothétique (âge > 70 ans, énolisme et déficits neuro-musculaires). Dans le groupe de 140 PTH à couple de frottement céramique-pe, le taux de luxation était de 6,2 % (malgré l insert en PE à débord postérieur utilisé) alors qu il chutait à 0,9 % dans le groupe de 109 PTH identiques à couple de frottement métal-métal, cette différence étant significative (p = 0,02). Les auteurs cherchèrent à expliquer ce résultat en menant une étude expérimentale sur les effets combinés du moment d extraction et de la cinétique du mouvement nécessaires à l obtention de la décoaptation entre les éléments des couples de frottement. Pour toutes les combinaisons étudiées, la force de rétention s opposant à la séparation des éléments du couple de frottement s est montrée significativement supérieure (ANOVA, p < 0,001) pour le couple CoCr-CoCr (maximum, 30 N à 10 cm/s) à celle relevée pour le couple céramique-pe (maximum, 4,1 N à 50 cm/s). Or, la même différence existait dans notre série prospective, mais non randomisée, de 340 PTH primaires non cimentées Alloclassic (tiges SL et anneaux vissés CSF) réalisées consécutivement de mai 1995 à octobre 2003, par le même opérateur et par la même voie d abord postéro-latérale que celle utilisée par Clarke et al. (15). Nous avons colligé un taux de luxations de 8,2 % pour le groupe de 170 PTH à couple de frottement alumine-pe- 28 mm et de 2,3 % avec le couple Metasul-28 mm, différence elle aussi significative (Chi 2 avec correction de continuité, p = 0,03). L âge des patients de part et d autre de 65 ans n avait aucun effet significatif sur cette différence (p = 0,59). À l origine de cette meilleure cohésion entre les composants métalliques du couple de frottement, l hypothèse d un effet «d adhésivité et de succion» établi sous l influence de la dureté du matériel (CoCr), de la viscosité du film liquidien et du jeu articulaire inter-prothétique (clearance) a été avancée. Un tel effet, bien illustré par la figure 3,

5 74 C. DELAUNAY FIG.3. Illustration du phénomène de «suction fit» maintenant la cohésion contre pesanteur d un couple métal-métal 28 mm (a) par rapport à un couple céramique d alumine-polyéthylène 28 mm (b) [extrait de l article de Clarke et al. (15), avec permission]. a b pourrait en partie expliquer la réduction de la fréquence des luxations observé par Clarke et al. (15) et par nous-mêmes. Ces informations sont à interpréter à la lumière des travaux expérimentaux récents menés par les équipes de Nassut et al. (16) et Nevelos et al. (17) rapportant les effets extrêmement péjoratifs sur l usure des couples de frottement de l introduction dans le fonctionnement des simulateurs de hanche des phénomènes d arrêt, de redémarrage et de décoaptation («piston»). Ces phénomènes tentent de simuler plus étroitement les modifications des rapports inter-prothétiques qui surviennent entre les deux composants du couple de frottement lors de mouvements de la vie courante, en particulier la décoaptation (sliding) bien mise en évidence in vivo par Lombardi et al. (18) lors de l enregistrement de la marche sur tapis roulant à l aide de la vidéo-fluoroscopie et après traitement informatique permettant l acquisition d images 3D (fig. 4). Tout récemment, cette dernière équipe a comparé la cinétique in vivo de 2 groupes de 10 patients chacun prothésés par des PTH identiques excepté la nature de leur couple de frottement en 28 millimètres, CoCr-PE versus Cocr-CoCr «Ultima» [Komistek et al. (19)]. La décoaptation moyenne mesurée était de 2 mm (de 0,8 à 3,1 mm) pour les couples CoCr-PE et de 0,38 mm seulement (de 0,3 à 0,5 mm) pour les couples CoCr-CoCr, cette différence étant significative (test t de Student, p = 0,0021). La synthèse des ces données cliniques et tribologiques permet de conclure que, pour des combinaisons de même calibre, la stabilité entre les composants des articulations prothétiques CoCr-CoCr est supérieure à celle des couples comportant du PE conventionnel en frottement. DISCUSSION FIG.4. Principe de mesure de la décoaptation entre bille et insert en radio-cinéma [extrait de l article de Komistek et al. (19), avec permission]. Les arguments précédents sont à confronter aux informations publiées depuis 2000 sur les autres couples de frottement. Alors que cimentée ou non, la fixation acétabulaire des implants Metasul ne pose toujours pas de problème particulier, ce que confirme la série de Long et al. (13), l inverse a été observé depuis 30 ans pour la fixation des cupules avec noyau d alumine massive, en particulier en France. L échec des fixations cimentées et non cimentées par anneau vissé ou impaction directe (Cerapress ), vient encore d être tout récemment confirmée par l équipe de Lariboisière dans une série d ostéonécroses de hanche, avec

6 UN COUPLE DE FROTTEMENT MÉTAL-MÉTAL PEUT-IL AMÉLIORER LA LONGÉVITÉ DES PROTHÈSES TOTALES DE HANCHE? % seulement de survie à 15 ans pour l événement «révision acétabulaire pour descellement aseptique» [Nich et al. (20)]. Dans le même ordre d idée, les résultats de la série originale de PTH à couple alumine-alumine de Boutin, indiquaient une survie des implants acétabulaires de 61,2 % seulement à 20 ans [Hamadouche et al. (21)]. Entre temps, la cupule non-cimentée à treillis de titane Cerafit (Ceraver-Osteal, Roissy), qui était présentée depuis 1990 comme l option de choix pour la fixation acétabulaire du couple alumine-alumine, est actuellement remplacée par la cupule Cerafit-HA dont les premières implantations ont débuté en 1998 et n ont donc que 6 ans de recul. En 2003 enfin, la même équipe faisait le point sur les 13 fractures d implants en alumine connues parmi leurs propres implants : 8 fractures de bille sur (dont en face d une cupule en PE) et 5 fractures d insert sur 3 300, soit un risque fracturaire globalement de l ordre de 15 pour [Hannouche et al. (22)]. Quant à l option céramique d alumine acétabulaire avec PE «en sandwich», le risque fracturaire évoqué par l auteur en 2000 s est vu confirmé depuis, non pas tant en France, avec quelques déclarations à l AFSSAPS, mais surtout au Japon, où la société Kyocera Inc. était depuis 1995 chef de file de cette option en 28 millimètres, avec un taux de fracture élevé de l ordre de 0,6 % (34 cas sur > inserts ABS ) rapporté par Hasegawa et al. (23). En ce qui concerne les nouveaux PE hautement réticulés, de nombreuses études cliniques sont actuellement en cours de réalisation. Signalons tout particulièrement l étude randomisée de l université de Göteborg dont les premiers résultats viennent d être communiqués par Digas et al. (24). À deux ans et sur 61 hanches, la radio-stéréométrie (RSA) et la radio-absorbtiométrie (DEXA) ne mettaient en évidence aucune différence significative entre les cupules de Weber cimentées (au Palacos-gentamycine) réalisées en PE conventionnel (Sulene, Zimmer-Centerpulse) ou en PE hautement réticulé (Durasul, Zimmer-Centerpulse). Il est intéressant de noter que sur les 2 premières années d usage in vivo chez des patients relativement jeunes (55 ans d age moyen) et normalement actifs, l usure moyenne tri-dimensionnelle globale était de 0,22 mm pour le PE-conventionnel et déjà de 0,18 mm pour le PE-Durasul (différence non significative, p = 0,1). Ce dernier chiffre de l ordre de 0,09 mm d usure moyenne linéaire annuelle, se situe dans l ordre de grandeur des données plus anciennement rapportées dans de nombreuses études en 22,2 ou 28 mm avec des couples de frottement standards [Delaunay (4)]. Avec un autre matériel hautement réticulé (PE Marathon, De Puy, Warsaw, États-Unis), à plus de 2 ans de recul moyen, la réduction d usure était de 81 % par rapport à un PE conventionnel, de l ordre de 0,02 millimètres linéaires par an, soit 17 mm 3 [Heisel et al. (25)]. Ces chiffres plus satisfaisants restent cependant en contradiction avec les données tribologiques obtenues in vitro selon lesquelles aucune usure n était détectable en 22 et 28 mm après 20 millions de cycles sur simulateur, ce qui correspond à environ 10 années d usage in vivo [Muratoglu et al. (26)]. Cette discordance entre les données d usure des implants en PE obtenues in vitro et in vivo n est pas nouvelle et semble devoir se répéter avec les PE à haute réticulation, du moins au court recul des 2 premières années. Elle est évoquée par l analyse de 21 explants d un PE hautement réticulé, montrant d importantes lésions de surface (abrasions, délamination, rayures) qui n avaient jamais été rapportées dans les études sur simulateur [Bradford et al. (27)]. Des réserves sont néanmoins encore d actualité concernant les couples de frottement métal-métal de type Metasul, en particulier en ce qui concerne les risques toxiques potentiels liés à la dissémination prolongée des débris métalliques dans la circulation générale ou à leur possible accumulation à distance dans l organisme [Jacobs et al. (28)]. En ce qui concerne le risque hématologique, il ne paraît plus aussi immédiat, la mortalité par cancer étant de 27,5 pour personnes-années dans le groupe des patients porteurs d une PTH métal-métal de première génération (McKee-Farrar) pour 39/ chez les patients porteurs de PTH couple de frottement conventionnels métal- PE dans l étude finlandaise de Visuri et Pukkala (29). Tout récemment, Visuri lors d une communication orale à la réunion de l EFORT (Helsinki, 2003) a rapporté, dans une population de 600 patients porteurs de PTH à couples métal-métal «modernes», une fréquence de cancer «normalement» attendue, mais sur une période de surveillance de 8 années seulement (données non publiées). Ce recul est certainement insuffisant pour être définitivement rassurant sur ce point. L élévation de la cobalthémie dans le sérum ou le sang total des patients porteurs d une prothèse de hanche avec un couple de frottement Metasul est une donnée bien établie depuis les travaux de Brodner et al. (30). Ce type d information peut être interprétée de 2 façons opposées : négativement, si l on considère que toute élévation anormale (augmentation de la production ou réduction des capacités d excrétion par insuffisance rénale, par exemple) sera la source de manifestations potentiellement «toxiques» à plus ou moins long terme : positivement, si l on considère que les débris d usure des couples de frottement métalliques sont «naturellement» éliminables et ne s accumulent donc pas dans l effective joint space, ce qui ne peut avoir qu un effet favorable sur la survenue d une ostéolyse périprothétique. De plus, la surveillance du taux de cobalt circulant pourrait représenter un excellent moyen de surveillance du fonctionnement intime du couple de frottement. Dans l expérience de l auteur, cette surveillance biologique s est montrée efficace, permettant de dépister certains dysfonctionnement mécaniques à type d effets came col-cupule (impingement), tout particulièrement mal tolérés par les couples «dur-sur-dur» (alumine-alumine et métal-métal). Dans l étude portant sur les 100 premières PTH Metasul de l auteur, ces effets cames ont été détectés suite à des élévations anormales de la cobalthémie qui étaient significative-

7 76 C. DELAUNAY ment liées à des antéversions exagérées des cupules acétabulaires (p = 0,0001) et à la présence de billes à jupe du premier modèle Metasul commercialisé de 1988 à 1995 [Delaunay (31)]. Antéversion exagérée de l anneau vissé et effet came lié à une bille à jupe ont généré le taux de cobalt le plus élevé de la série (36 g/l de sang total) et ont conduit à la nécessité de révision de l implant fémoral correspondant, secondairement désancré après 7 ans et 8 mois de fonctionnement in vivo. Cette révision a été immédiatement suivie d une chute de la cobalthémie, à 2,5 g/l au 2 e mois et indétectable au dernier contrôle. En revanche, l absence de relation significative entre l élévation de la cobalthémie et le niveau d activité le plus élevé de nos patients (Devane 4 et 5 pour 33 % des cas avec un âge moyen de 59 ans pour l ensemble de la série) représente une information particulièrement positive dans la mesure où ce type de couple de frottement à haute résistance à l usure est justement avant tout destiné aux patients les plus actifs, qui ne sont d ailleurs pas forcément les plus jeunes. Quant au retour à des arthroplasties de resurfaçage avec des implants métalmétal de grande taille, les premiers résultats clinique publiés sont prometteurs [Daniel et al. (32), Amstutz et al. (33)]. Cependant, dans l étude comparative de Clarke et al. (34), ces grands calibres s accompagnent d une élévation moyenne du cobalt sérique plus importante, de l ordre de 38 nmol/l versus 22 nmol/l pour le calibre 28 millimètres plus conventionnel. Néanmoins, de nombreuses inconnues persistent encore sur les réelles performances tribologiques des couples métal-métal disponibles sur le marché, mais dont les caractéristiques métallurgiques ne sont pas identiques. Dans une mise au point récente, Nevelos et al. (35) confirment que les performances des couples métal-métal forgés à haute concentration de carbone pour les 2 composants, ce qui est le cas du Metasul, sont bonnes dans les circonstances de fonctionnement physiologiques, alors que la résistance à l usure des couples métal-métal à basse concentration de carbone pour les 2 composants n est pas satisfaisante, ce qui avait été observé cliniquement avec le couple Sikomed SM21 (EndoPlus) abandonné depuis [(Delaunay (4)]. Il est surtout important de resituer la production de débris métalliques provenant d un couple de frottement prothétique au sein de toutes les autres sources possiblement induites par l introduction de corps étrangers métalliques dans l organisme humain. Des élévations systémiques du cobalt et du chrome ont été aussi mises en évidence avec des PTH à couple de frottement «conventionnel» métal-pe descellées, non cimentées [Kreibich et al. (36)] ou cimentées [Damie et Favard (37)], mais aussi non descellées cimentées [Harding et al. (38)], comme avec des prothèses totales de genou non cimentées [Liu et al. (39)], des clous centro-médullaires modulaires [Jones et al. (40)], etc. À l évidence, le recours à un couple de frottement métal-métal ne représente qu une origine de plus au sein de toutes les sources possibles de débris métalliques circulant. Reste le problème des réactions d hypersensibilité retardée mises en évidence par Willert et al. (41) dans les tissus recueillis autour de PTH à couples de frottement métalmétal lors de leur révision. La fréquence des allergies aux métaux, de l ordre de 10 à 15 % dans la population générale, double chez les patients porteurs d implants métalliques, et quintuple en cas d échec (descellement d arthroplastie) ou de dysfonctionnement pour atteindre 50 à 60 % [Jacobs et al. (28)]. La question, toujours en suspens, est de savoir si l allergie métallique est la cause du descellement ou sa conséquence. Dans le cas de révision fémorale pour descellement rapporté par l auteur, il existait un dysfonctionnement mécanique évident par effet came entre la jupe de la bille et le débord de la cupule acétabulaire Metasul du premier modèle, mais l analyse histologique des tissus, réalisée par l équipe de Willert, retrouvait aussi des réactions cellulaires indiquant un processus d hypersensibilité retardée de type IV. Il est tout aussi inconnu à ce jour de savoir si l allergie métallique est un phénomène sporadique ne concernant que quelques patients «susceptibles», où s il s agit d un phénomène beaucoup plus courant jouant un rôle important dans la genèse des échecs des implants. Ces considérations sont d une importance primordiale pour les patients porteurs de PTH à couple de frottement métalmétal présentant une élévation chronique des concentrations systémiques d ions métalliques. CONCLUSION Avec le couple de frottement Metasul ont été observées radiographiquement, une réduction de la migration précoce des implants acétabulaires ainsi que de l usure et de la survenue de l ostéolyse à court terme, et cliniquement, une réduction de la fréquence des luxations. Ainsi, après seulement huit années d expérience en France, et dans l état actuel des données de la littérature et des observations cliniques en cours, le couple de frottement CoCr-sur-CoCr Metasul représente bien encore, en parallèle avec les couples alumine-alumine et le recours aux nouveaux polyéthylènes hautement réticulés, une option potentiellement susceptible d améliorer la longévité des arthroplasties totales de la hanche. Les réponses aux questions non résolues sur la carcinogenèse et surtout sur l importance du risque immuno-allergique ne seront obtenues que lorsque l on disposera d études épidémiologiques multicentriques à long terme et de meilleures connaissances en biologie moléculaire et dans la science des bio-matériaux. Références 1. MALCHAU H, HERBERTS P, AHNFELT L : Prognosis of total hip replacement in Sweden. Follow-up of 92,675 operations performed Acta Orthop Scand, 1993, 64,

8 UN COUPLE DE FROTTEMENT MÉTAL-MÉTAL PEUT-IL AMÉLIORER LA LONGÉVITÉ DES PROTHÈSES TOTALES DE HANCHE? MALCHAU H, HERBERTS P, EISLET T, GARELLICK G, SÖDERMAN P : The Swedish total hip replacement Register. J Bone Joint Surg (Am), 2002, 84 (suppl 2), DELAUNAY C : Couple de friction métallique de seconde génération en arthroplastie totale primaire non cimentée de hanche. Justification, homologation française et résultats préliminaires. Rev Chir Orthop, 2000, 86, DELAUNAY C : Couple de frottement des prothèses totales de hanche. Ce qu un chirurgien orthopédiste devrait savoir! Conférences d enseignement. Cahiers d enseignement de la SOFCOT, 2001, 78, HEISEL C, SILVA M, SCHMALZRIED TP : Bearing surface options for total hip replacement in young patients. J Bone Joint Surg (Am), 2003, 85, STOCKS GW, FREEMAN MAR, EVANS SJW : Acetabular cup migration: prediction of aseptic loosening. J Bone Joint Surg (Br), 1995, 77, ILCHMANN T, KESTERIS U, WINGSTRAND H : EBRA improves the accuracy of radiographic analysis of acetabular cup migration. Acta Orthop Scand, 1998, 69, BOEHLER M, PLENK H, SALZER M : Alumina ceramic bearings for hip endoprostheses: the Austrian experiences. Clin Orthop, 2000, 379, PABINGER C, BIEDERMANN R, STÖCKL B, FISCHER M, KRISMER M : Migration of metal-on-metal versus ceramicon-polyethylene hip prostheses. Clin Orthop, 2003, 412, DELAUNAY C, KAPANDJI AI : Survival analysis of cementless grit-blasted titanium total hip arthroplasties. J Bone Joint Surg (Br), 2001, 83, DELAUNAY C, BONNOMET F, NORTH J, JOBARD D, CAZEAU C, KEMPF JF : Grit-blasted titanium femoral stem in cementless primary total hip arthroplasty A 5- to 10-year multicenter study. J Arthroplasty, 2001, 16, BONNOMET F, DELAUNAY C, SIMON P, LEFEBVRE Y, CLAVERT P, KAPANDJI AI et al. : Comportement d une tige fémorale droite en arthroplastie totale primaire non cimentée de la hanche chez les patients de moins de 65 ans. Rev Chir Orthop, 2001, 87, LONG WT, DORR LD, GENDELMAN V : An american experience with metal-on-metal total hip arthroplasties : a 7-year follow-up study. J Arthroplasy, 2004, 19 (suppl 3), MIGAUD H, JOBIN A, LAFFARGUE P, GIRAUD F, DUQUENNOY A : Cementless metal on metal hip replacement in patient less than fifty years of age: comparison with a watched control group using ceramic on polyethylene after a minimum five-year follow-up. J Arthroplasty, 2004, 19 (suppl 3), CLARKE MT, LEE PTH, VILLAR RN : Dislocation after total hip replacement in relation to metal-on-metal bearing surfaces. J Bone Joint Surg (Br), 2003, 85, NASSUT R, WIMMER MA, SCHNEIDER E, MORLOCK MM : The influence of resting periods on friction in the artificial hip. Clin Orthop, 2003, 407, NEVELOS J, INGHAM E, DOYLE C, STREICHER R, NEVELOS A, WALTER W et al. : Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty, 2000, 15, LOMBARDI AV JR, MALLORY TH, DENNIS DA, KOMISTEK RD, FADA RA, NORTHCUT EJ : An in vivo determination of total hip arthroplasty pistoning during activity. J Arthroplasty, 2000, 15, KOMISTEK RD, DENNIS DA, OCHOA JA, HAAS BD, HAMMIL C: In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg (Am), 2002, 84, NICH C, EL-HADI SARI A, HANNOUCHE D, NIZARD R, WITVOET J, SEDEL L et al. : Long-term results of Aluminaon-Alumina hip arthroplasty for osteonecrosis. Clin Orthop, 2003, 417, HAMADOUCHE M, BOUTIN P, DAUSSANGE J, BOLANDER M, SEDEL L : Alumina-on-Alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg (Am), 2002, 84, HANNOUCHE D, NICH C, BIZOT P, MEUNIER A, NIZARD R, SEDEL L : Fracture of ceramic bearings. History and present status. Clin Orthop, 2003, 417, HASEGAWA M, SUDO A, HIRATA H, UCHIDA A : Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty, 2003, 18, DIGAS G, KÄRRHOLM J, THANNER J, MALCHAU H, HERBERTS P : Highly cross-linked polyethylene in cemented THA. Randomized study of 61 hips. Clin Orthop, 2003, 417, HEISEL C, SILVA M, DELA ROSA A, SCHMALZRIED TP : Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg (Am), 2004, 86, MURATOGLU OK, BRAGDON CR, O CONNOR DO, JASTY M, HARRIS WH : A novel method of cross-linking ultra-highmolecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. J Arthroplasty, 2001, 16, BRADFORD L, BAKER DA, GRAHAM J, CHAWAN A, RIES MD, PRUITT LA : Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners. J Bone Joint Surg (Am), 2004, 6, JACOBS JJ, HALLAB NJ, SKIPOR AK, URBAN RM : Metal degradation products. A cause for concern in Metal-Metal bearings? Clin Orthop, 2003, 417, VISURI T, PUKKALA E : Does Metal-on-Metal total hip prosthesis have influence on cancer? In : RIEKER C, OBERHOLZER S, WYSS U (eds): World tribology forum in arthroplasty. Huber, Bern, 2001, p BRODNER W, BITZAN P, MEISINGER V, KAIDER A, GOTTSAUNER-WOLF F, KOTZ R : Elevated serum cobalt with metal-on-metal articulating surfaces. J Bone Joint Surg (Am), 2003, 85, DELAUNAY C : Metal-on-metal bearings in primary total hip arthroplasty. A personal series of 100 consecutive cementless alloclassic-metasul hips in France. J Arthroplasty, 2004, 19, (suppl 3). 32. DANIEL J, PYNSENT PB, MCMINN DJW : Metal-on-metal resurfacing of the hip in patients under the age of 55 years with osteoarthritis. J Bone Joint Surg (Br), 2004, 86, AMSTUTZ HC, BAULE PE, DOREY FJ, LE DUFF MJ, CAMPBELL PA, GRUEN TA : Metal-on-metal hybrid surface

9 78 C. DELAUNAY arthroplasty: two to six-year follow-up study. J Bone Joint Surg (Am), 2004, 86, CLARKE MT, LEE PTH, ARORA A, VILLAR RN : Levels of metal ions after small- and large-diameter metal-onmetal hip arthroplasty. J Bone Joint Surg (Br), 2003, 85, NEVELOS J, SHELTON JC, FISHER J : Metallurgical considerations in the wear of metal-on-metal hip bearings. Hip International, 2004, 14, KREIBICH DN, MORAN CG, DELVES HT, OWEN TD, PINDER IM : Systemic release of cobalt and chromium after uncemented total hip replacement. J Bone Joint Surg (Br), 1996, 78, DAMIE F, FAVARD L : Dosages sanguins de métaux chez 48 patients porteurs d une prothèse totale de hanche en chrome-cobalt à couple métal-polyéthylène. Rev Chir Orthop, 2004, 90, HARDING I, BONOMO A, CRAWFORD R, PSYCHOYIOS V, DELVES T, MURRAY D et al. : Serum levels of cobalt and chromium in a complex modular total hip arthroplasty system. J Arthroplasty, 2002, 17, LIU TK, LIU SH, CHANG CH, YANG RS : Concentration of metal elements in the blood and urine in the patients with cementless total knee arthroplasty. Tohoku J Exp Med, 1998, 185, JONES DM, MARSH JL, NEPOLA JV, JACOBS JJ, SKIPOR AK, URBAN RM et al. : Focal osteolysis at the junctions of a modular stainless-steel femoral intramedullary nail. J Bone Joint Surg (Am), 2001, 83, WILLERT HG, BUCHHORN GH, FAYYAZI A, LOHMANN CH : Histopathological changes in tissues surrounding metal/metal joints Signs of delayed type hypersensitivity (DTH). In : RIEKER C, OBERHOLZER S, WYSS U (eds): World tribology forum in arthroplasty. Huber, Bern, 2001, p

J.H. CATON clinique E. De Vialar - Lyon, J.L. PRUDHON Clinique des Cèdres - Echirolles A. FERREIRA Clinique du Parc Lyon T. ASLANIAN Groupe Lépine -

J.H. CATON clinique E. De Vialar - Lyon, J.L. PRUDHON Clinique des Cèdres - Echirolles A. FERREIRA Clinique du Parc Lyon T. ASLANIAN Groupe Lépine - J.H. CATON clinique E. De Vialar - Lyon, J.L. PRUDHON Clinique des Cèdres - Echirolles A. FERREIRA Clinique du Parc Lyon T. ASLANIAN Groupe Lépine - Genay J.H. CATON Consultant et royalties Groupe Lépine

Plus en détail

MATTHIEU GUYARD SEBASTIEN MARTRES GUALTER VAZ JEAN-PAUL CARRET

MATTHIEU GUYARD SEBASTIEN MARTRES GUALTER VAZ JEAN-PAUL CARRET MATTHIEU GUYARD SEBASTIEN MARTRES GUALTER VAZ JEAN-PAUL CARRET PROTHESE TOTALE DE HANCHE POUR OSTEONECROSE ASEPTIQUE DE LA TETE FEMORALE Analyse d une série de 126 arthroplasties. Hôpital Edouard HERRIOT

Plus en détail

CONTROVERSES en chirurgie de la hanche. C. Schwartz Conflit d intérêt: FH Orthopedics Colmar

CONTROVERSES en chirurgie de la hanche. C. Schwartz Conflit d intérêt: FH Orthopedics Colmar CONTROVERSES en chirurgie de la hanche C. Schwartz Conflit d intérêt: FH Orthopedics Colmar Une invention française Gilles Bousquet CHU de Saint-Etienne - Jean Rieu, Professeur à l Ecole des Mines de Saint-Etienne

Plus en détail

LES NOUVEAUX ANTICOAGULANTS

LES NOUVEAUX ANTICOAGULANTS LES NOUVEAUX ANTICOAGULANTS Pr. Alessandra Bura-Rivière, Service de Médecine Vasculaire Hôpital Rangueil, 1 avenue Jean Poulhès, 31059 Toulouse cedex 9 INTRODUCTION Depuis plus de cinquante ans, les héparines

Plus en détail

Une avancée majeure dans le domaine de l implantologie. Roxolid SLActive Moins invasif pour de nouveaux standards

Une avancée majeure dans le domaine de l implantologie. Roxolid SLActive Moins invasif pour de nouveaux standards Une avancée majeure dans le domaine de l implantologie. Roxolid SLActive Moins invasif pour de nouveaux standards 1 Excellentes performances scientifiquement démontrées. Roxolid est un matériau unique

Plus en détail

Intérêt diagnostic du dosage de la CRP et de la leucocyte-estérase dans le liquide articulaire d une prothèse de genou infectée

Intérêt diagnostic du dosage de la CRP et de la leucocyte-estérase dans le liquide articulaire d une prothèse de genou infectée Intérêt diagnostic du dosage de la CRP et de la leucocyte-estérase dans le liquide articulaire d une prothèse de genou infectée C. Rondé-Oustau, JY. Jenny,J.Sibilia, J. Gaudias, C. Boéri, M. Antoni Hôpitaux

Plus en détail

L. Obert, T. Lascar, A. Adam

L. Obert, T. Lascar, A. Adam Améliorer la consolidation des tubérosités grâce au système OMS (Offset Modular System) et l autogreffe Improving tuberosity consolidation using the OMS system (Offset Modular System) L. Obert, T. Lascar,

Plus en détail

Qui et quand opérer. au cours du traitement de l EI?

Qui et quand opérer. au cours du traitement de l EI? Qui et quand opérer au cours du traitement de l EI? Gilbert Habib Département de Cardiologie - Timone Marseille 7es JNI Bordeaux, 8 juin 2006 Université de la Méditerranée Faculté de Médecine de Marseille

Plus en détail

ACCIDENTS ELECTRIQUES EN CHIRURGIE COELIOSCOPIQUE. Dr JF Gravié FCVD

ACCIDENTS ELECTRIQUES EN CHIRURGIE COELIOSCOPIQUE. Dr JF Gravié FCVD ACCIDENTS ELECTRIQUES EN CHIRURGIE COELIOSCOPIQUE Dr JF Gravié FCVD CONTEXTE (1) Risques électriques de 0,1 à 0,4% Enquête de pratique * 506 chirurgiens: 18% expérience personnelle, 54% connaissaient un

Plus en détail

ÉVALUATION DE L UTILISABILITÉ D UN SITE WEB : TESTS D UTILISABILITÉ VERSUS ÉVALUATION HEURISTIQUE

ÉVALUATION DE L UTILISABILITÉ D UN SITE WEB : TESTS D UTILISABILITÉ VERSUS ÉVALUATION HEURISTIQUE ÉVALUATION DE L UTILISABILITÉ D UN SITE WEB : TESTS D UTILISABILITÉ VERSUS ÉVALUATION HEURISTIQUE BOUTIN MARIO Centre de recherche informatique de Montréal (CRIM), 550, rue Sherbrooke Ouest, Bureau 100,

Plus en détail

Chirurgie Mini-Invasive par Voie Antérieure La PTH sans section musculaire

Chirurgie Mini-Invasive par Voie Antérieure La PTH sans section musculaire Chirurgie Mini-Invasive par Voie Antérieure La PTH sans section musculaire Clinique Saint-George Nice Dr Didier D Hondt Chirurgien orthopédiste Vincent Christian Kinésithérapeute Jeudi 20 Mars 2014 Nice

Plus en détail

e-santé du transplanté rénal : la télémédecine au service du greffé

e-santé du transplanté rénal : la télémédecine au service du greffé e-santé du transplanté rénal : la télémédecine au service du greffé Professeur Michèle Kessler CHU de Nancy et réseau Néphrolor L une des applications de la télémédecine est la télésurveillance à domicile,

Plus en détail

Gelsoft Plus LA SEULE PROTHESE VASCULAIRE AU MONDE EN POLYESTER A STRUCTURE TRICOTEE KOPER IMPREGNEE DE GELATINE, RESISTANT A LA DILATATION

Gelsoft Plus LA SEULE PROTHESE VASCULAIRE AU MONDE EN POLYESTER A STRUCTURE TRICOTEE KOPER IMPREGNEE DE GELATINE, RESISTANT A LA DILATATION Gelsoft Plus LA SEULE PROTHESE VASCULAIRE AU MONDE EN POLYESTER A STRUCTURE TRICOTEE KOPER IMPREGNEE DE GELATINE, RESISTANT A LA DILATATION Structure tricotée Köper exclusive offrant: - une dilatation

Plus en détail

La reprise de la vie active

La reprise de la vie active Prothèse de hanche Les conseils de votre chirurgien La reprise de la vie active Vous êtes aujourd hui porteur d une prothèse de hanche ; elle va vous permettre de reprendre vos activités sans gène ni douleur.

Plus en détail

QUEL AVENIR POUR LA PHARMACIE HOSPITALIERE EN SUISSE?

QUEL AVENIR POUR LA PHARMACIE HOSPITALIERE EN SUISSE? QUEL AVENIR POUR LA PHARMACIE HOSPITALIERE EN SUISSE? Regard critique sur nos filières de formation Symposium GSASA, Berne, 12 mai 2009 André Pannatier 1 PLAN 1. Les filières de formation de base et postdiplôme

Plus en détail

UNIVERSITY OF MALTA FACULTY OF ARTS. French as Main Area in an ordinary Bachelor s Degree

UNIVERSITY OF MALTA FACULTY OF ARTS. French as Main Area in an ordinary Bachelor s Degree French Programme of Studies (for courses commencing October 2009 and later) YEAR ONE (2009/10) Year (These units start in and continue in.) FRE1001 Linguistique théorique 1 4 credits Non Compensatable

Plus en détail

INFORMATION À DESTINATION DES PROFESSIONNELS DE SANTÉ LE DON DU VIVANT

INFORMATION À DESTINATION DES PROFESSIONNELS DE SANTÉ LE DON DU VIVANT INFORMATION À DESTINATION DES PROFESSIONNELS DE SANTÉ LE DON DU VIVANT QUELS RÉSULTATS POUR LE RECEVEUR? QUELS RISQUES POUR LE DONNEUR? DONNER UN REIN DE SON VIVANT PEUT CONCERNER CHACUN /////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

Plus en détail

Essais précoces non comparatifs : principes et calcul du nombre de sujets nécessaire

Essais précoces non comparatifs : principes et calcul du nombre de sujets nécessaire Essais précoces non comparatifs : principes et calcul du nombre de sujets nécessaire Sylvie CHABAUD Direction de la Recherche Clinique et de l Innovation : Centre Léon Bérard - Lyon Unité de Biostatistique

Plus en détail

Activité 38 : Découvrir comment certains déchets issus de fonctionnement des organes sont éliminés de l organisme

Activité 38 : Découvrir comment certains déchets issus de fonctionnement des organes sont éliminés de l organisme Activité 38 : Découvrir comment certains déchets issus de fonctionnement des organes sont éliminés de l organisme 1. EXTRAITS REFERENTIELS DU BO Partie du programme : Fonctionnement de l organisme et besoin

Plus en détail

Quel apport de l imagerie dans les traitements anti-angiogéniques?

Quel apport de l imagerie dans les traitements anti-angiogéniques? Quel apport de l imagerie dans les traitements anti-angiogéniques? L Fournier, D Balvay, CA Cuénod Service de radiologie, Hôpital Européen Georges Pompidou Laboratoire de Recherche en Imagerie, Equipe

Plus en détail

Revue de la littérature

Revue de la littérature Résultats à long terme des anévrysmes traités Revue de la littérature Cours de DES de neurochirurgie Inter-région Rhône-Alpes / Auvergne Vendredi 22/04/11 Méthode (1): sélection bibliographique en fonction

Plus en détail

Laurence LEGOUT, Michel VALETTE, Henri MIGAUD, Luc DUBREUIL, Yazdan YAZDANPANAH et Eric SENNEVILLE

Laurence LEGOUT, Michel VALETTE, Henri MIGAUD, Luc DUBREUIL, Yazdan YAZDANPANAH et Eric SENNEVILLE Laurence LEGOUT, Michel VALETTE, Henri MIGAUD, Luc DUBREUIL, Yazdan YAZDANPANAH et Eric SENNEVILLE Service Régional Universitaires des Maladies Infectieuses et du Voyageur, Hôpital Gustave Dron 59208 TOURCOING

Plus en détail

Item 169 : Évaluation thérapeutique et niveau de preuve

Item 169 : Évaluation thérapeutique et niveau de preuve Item 169 : Évaluation thérapeutique et niveau de preuve COFER, Collège Français des Enseignants en Rhumatologie Date de création du document 2010-2011 Table des matières ENC :...3 SPECIFIQUE :...3 I Différentes

Plus en détail

Validation clinique des marqueurs prédictifs le point de vue du méthodologiste. Michel Cucherat UMR CNRS 5558 - Lyon

Validation clinique des marqueurs prédictifs le point de vue du méthodologiste. Michel Cucherat UMR CNRS 5558 - Lyon Validation clinique des marqueurs prédictifs le point de vue du méthodologiste Michel Cucherat UMR CNRS 5558 - Lyon Marqueur prédictif - Définition Un marqueur prédictif est un marqueur qui prédit le bénéfice

Plus en détail

Réflexions sur les possibilités de réponse aux demandes des chirurgiens orthopédistes avant arthroplastie

Réflexions sur les possibilités de réponse aux demandes des chirurgiens orthopédistes avant arthroplastie Bull. Acad. Natle Chir. Dent., 2007, 50 113 Commission de l exercice professionnel et Groupe de réflexion Réflexions sur les possibilités de réponse aux demandes des chirurgiens orthopédistes avant arthroplastie

Plus en détail

EVALUATION DES TECHNOLOGIES DE SANTÉ ANALYSE MÉDICO-ÉCONOMIQUE. Efficacité et efficience des hypolipémiants Une analyse centrée sur les statines

EVALUATION DES TECHNOLOGIES DE SANTÉ ANALYSE MÉDICO-ÉCONOMIQUE. Efficacité et efficience des hypolipémiants Une analyse centrée sur les statines EVALUATION DES TECHNOLOGIES DE SANTÉ ANALYSE MÉDICO-ÉCONOMIQUE Efficacité et efficience des hypolipémiants Une analyse centrée sur les statines Juillet 2010 Mise à jour Septembre 2010 1 Le rapport complet

Plus en détail

Controverse UDM télésurveillée Pour. P. Simon Association Nationale de Télémédecine

Controverse UDM télésurveillée Pour. P. Simon Association Nationale de Télémédecine Controverse UDM télésurveillée Pour P. Simon Association Nationale de Télémédecine Controverse Peut-on être en 2013 opposé au développement de la télémédecine? Pourquoi les patients en insuffisance rénale

Plus en détail

TSTI 2D CH X : Exemples de lois à densité 1

TSTI 2D CH X : Exemples de lois à densité 1 TSTI 2D CH X : Exemples de lois à densité I Loi uniforme sur ab ; ) Introduction Dans cette activité, on s intéresse à la modélisation du tirage au hasard d un nombre réel de l intervalle [0 ;], chacun

Plus en détail

LIGNES DIRECTRICES CLINIQUES TOUT AU LONG DU CONTINUUM DE SOINS : Objectif de ce chapitre. 6.1 Introduction 86

LIGNES DIRECTRICES CLINIQUES TOUT AU LONG DU CONTINUUM DE SOINS : Objectif de ce chapitre. 6.1 Introduction 86 LIGNES DIRECTRICES CLINIQUES TOUT AU LONG DU CONTINUUM DE SOINS : ÉTABLISSEMENT DE LIENS ENTRE LES PERSONNES CHEZ QUI UN DIAGNOSTIC D INFECTION À VIH A ÉTÉ POSÉ ET LES SERVICES DE SOINS ET DE TRAITEMENT

Plus en détail

TRAUMATISME CRANIEN DE L ENFANT : conduite à tenir?

TRAUMATISME CRANIEN DE L ENFANT : conduite à tenir? TRAUMATISME CRANIEN DE L ENFANT : conduite à tenir? H Le Hors-Albouze Urgences pédiatriques Timone enfants Marseille Traumatismes crâniens (TC) de l enfant Grande fréquence même si incidence réelle mal

Plus en détail

La recherche clinique au cœur du progrès thérapeutique

La recherche clinique au cœur du progrès thérapeutique 2 èmes Assises de l Innovation Thérapeutique Paris, 05-11-2014 La recherche clinique au cœur du progrès thérapeutique Pr. Jean-François DHAINAUT Président du Groupement Inter-régional de la Recherche Clinique

Plus en détail

Chirurgie assistée par robot et laparoscopie en 3D à l avantage des patients?

Chirurgie assistée par robot et laparoscopie en 3D à l avantage des patients? HigHligHts 2012: CHiRURgiE Chirurgie assistée par robot et laparoscopie en 3D à l avantage des patients? Lukasz Filip Grochola a, Hubert John b, Thomas Hess c, Stefan Breitenstein d a Clinique de chirurgie

Plus en détail

DENSITOMÉTRIE OSSEUSE : CE QUE LE RADIOLOGUE DOIT SAVOIR

DENSITOMÉTRIE OSSEUSE : CE QUE LE RADIOLOGUE DOIT SAVOIR DENSITOMÉTRIE OSSEUSE : CE QUE LE RADIOLOGUE DOIT SAVOIR C Andreux et F L Huillier H Guerini, A Feydy, X Poittevin, F Thevenin, R Campagna, JL Drapé, A Chevrot Hôpital COCHIN, Paris, France OBJECTIF Proposer

Plus en détail

Consensus Scientifique sur. les. Champs statiques

Consensus Scientifique sur. les. Champs statiques page 1/8 Consensus Scientifique sur Source : OMS (2006) les Champs statiques Résumé & Détails: GreenFacts Contexte - Les équipements d imagerie médicale par résonance magnétique (IRM), les trains et les

Plus en détail

Prix Pierre Potier L innovation en chimie au bénéfice de l environnement

Prix Pierre Potier L innovation en chimie au bénéfice de l environnement Prix Pierre Potier L innovation en chimie au bénéfice de l environnement A l initiative de François Loos Ministre délégué à l Industrie Page 1 Prix Pierre Potier L innovation en chimie au bénéfice de l

Plus en détail

Interest Rate for Customs Purposes Regulations. Règlement sur le taux d intérêt aux fins des douanes CONSOLIDATION CODIFICATION

Interest Rate for Customs Purposes Regulations. Règlement sur le taux d intérêt aux fins des douanes CONSOLIDATION CODIFICATION CANADA CONSOLIDATION CODIFICATION Interest Rate for Customs Purposes Regulations Règlement sur le taux d intérêt aux fins des douanes SOR/86-1121 DORS/86-1121 Current to August 4, 2015 À jour au 4 août

Plus en détail

FAQ CLEARFIL SE BOND. Pagina 1 van 5

FAQ CLEARFIL SE BOND. Pagina 1 van 5 FAQ CLEARFIL SE BOND 1. Quels sont les avantages des systèmes automordançants? 2. CLEARFIL SE BOND est-il le système adhésif le plus utilisé? 3. CLEARFIL SE BOND est-il le système adhésif le plus étudié?

Plus en détail

Pharmacovigilance des nouveaux anticoagulants oraux

Pharmacovigilance des nouveaux anticoagulants oraux Pharmacovigilance des nouveaux anticoagulants oraux Dr A.Lillo-Le Louët, Centre Régional de Pharmacovigilance (CRPV) Paris-HEGP Avec avec Dr P.Lainé, CRPV - Angers Préambule Directeur du Centre Régional

Plus en détail

Peut-on ne pas reprendre des marges «insuffisantes» en cas de Carcinome canalaire infiltrant

Peut-on ne pas reprendre des marges «insuffisantes» en cas de Carcinome canalaire infiltrant Peut-on ne pas reprendre des marges «insuffisantes» en cas de Carcinome canalaire infiltrant Institut Cancérologie de l Ouest CHIRURGIE Dr. Isabelle Jaffré CAS CLINIQUE 1 36 ans 90B sans CI radiothérapie

Plus en détail

ÉTUDE DE L EFFICACITÉ DE GÉOGRILLES POUR PRÉVENIR L EFFONDREMENT LOCAL D UNE CHAUSSÉE

ÉTUDE DE L EFFICACITÉ DE GÉOGRILLES POUR PRÉVENIR L EFFONDREMENT LOCAL D UNE CHAUSSÉE ÉTUDE DE L EFFICACITÉ DE GÉOGRILLES POUR PRÉVENIR L EFFONDREMENT LOCAL D UNE CHAUSSÉE ANALYSIS OF THE EFFICIENCY OF GEOGRIDS TO PREVENT A LOCAL COLLAPSE OF A ROAD Céline BOURDEAU et Daniel BILLAUX Itasca

Plus en détail

Malgré les évolutions technologiques. Prothèse implantaire fixe scellée et vissée à l heure du numérique. Spécial IMPLANTOLOGIE. prothèse implantaire

Malgré les évolutions technologiques. Prothèse implantaire fixe scellée et vissée à l heure du numérique. Spécial IMPLANTOLOGIE. prothèse implantaire prothèse implantaire Dans le cadre d une réhabilitation orale implantaire, la finalité de notre traitement est la prothèse qui est placée sur les implants. Pour que cette prothèse soit fonctionnelle et

Plus en détail

Donneurs vivants Risques à long terme. Cours de transplantation Univ. Montréal et McGill 5 avril 2013

Donneurs vivants Risques à long terme. Cours de transplantation Univ. Montréal et McGill 5 avril 2013 Donneurs vivants Risques à long terme Michel R. Pâquet MD, PhD Unité de Transplantation Le Centre Hospitalier de l Université de Montréal Hôpital Notre-Dame Montréal Hôtel-Dieu Hôpital Notre-Dame Cours

Plus en détail

Lavatory Faucet. Instruction Manual. Questions? 1-866-661-9606 customerservice@artikaworld.com

Lavatory Faucet. Instruction Manual. Questions? 1-866-661-9606 customerservice@artikaworld.com Lavatory Faucet Instruction Manual rev. 19-01-2015 Installation Manual You will need Adjustable Wrench Adjustable Pliers Plumber s Tape Hardware list (included) Allen Key Socket wrench tool Important Follow

Plus en détail

De meilleurs soins :

De meilleurs soins : De meilleurs soins : une analyse des soins infirmiers et des résultats du système de santé Série de rapports de l AIIC et de la FCRSS pour informer la Commission nationale d experts de l AIIC, La santé

Plus en détail

Programme «maladie» - Partie II «Objectifs / Résultats» Objectif n 2 : développer la prévention

Programme «maladie» - Partie II «Objectifs / Résultats» Objectif n 2 : développer la prévention Programme «maladie» - Partie II «Objectifs / Résultats» Objectif n 2 : développer la prévention Indicateur n 2-3 : Indicateurs sur le dépistage du cancer 1 er sous-indicateur : taux de participation au

Plus en détail

LA PROTHESE TOTALE DE GENOU

LA PROTHESE TOTALE DE GENOU LA PROTHESE TOTALE DE GENOU Introduction : La destruction du cartilage est fréquente au niveau du genou qui est une articulation très sollicitée et soumise à tout le poids du corps. La conséquence de cette

Plus en détail

CONSEILS POUR LA REDACTION DU RAPPORT DE RECHERCHE. Information importante : Ces conseils ne sont pas exhaustifs!

CONSEILS POUR LA REDACTION DU RAPPORT DE RECHERCHE. Information importante : Ces conseils ne sont pas exhaustifs! CONSEILS POUR LA REDACTION DU RAPPORT DE RECHERCHE Information importante : Ces conseils ne sont pas exhaustifs! Conseils généraux : Entre 25 et 60 pages (hormis références, annexes, résumé) Format d un

Plus en détail

Suivi Biologique des Nouveaux Anticoagulants

Suivi Biologique des Nouveaux Anticoagulants Suivi Biologique des Nouveaux Anticoagulants M.M. Samama, M-H. Horellou, C. Flaujac, J. Conard Groupe Hémostase-Thrombose Hôtel-Dieu L. Le Flem, C. Guinet, F. Depasse Biomnis - Ivry sur Seine TFPI TFPI

Plus en détail

Evaluer le risque fracturaire: l outil FRAX (Fracture Risk Assessment Tool)

Evaluer le risque fracturaire: l outil FRAX (Fracture Risk Assessment Tool) Evaluer le risque fracturaire: l outil FRAX (Fracture Risk Assessment Tool) F. FLAISLER Service de Rhumatologie Groupe Hospitalo-Universitaire CAREMEAU, Nîmes OSTÉOPOROSE DEFINITION AFSSAPS «Maladie diffuse

Plus en détail

CONTROVERSE : IDR OU QUANTIFERON LORS D'UN CONTAGE EN EHPAD?

CONTROVERSE : IDR OU QUANTIFERON LORS D'UN CONTAGE EN EHPAD? CONTROVERSE : IDR OU QUANTIFERON LORS D'UN CONTAGE EN EHPAD? Hélène MANGEARD François MALADRY Tuberculose : infection mycobactérienne Infection mycobactérienne chronique (M. Tuberculosis ++ ou bacille

Plus en détail

PREVENTION OF MSL FOR VDT USERS IN A LARGE SERVICE ORIENTED BUSINESS

PREVENTION OF MSL FOR VDT USERS IN A LARGE SERVICE ORIENTED BUSINESS INTERVENTION POUR PRÉVENIR LES TMS CHEZ LES UTILISATEURS DE TEV DANS UNE GRANDE ENTREPRISE DE SERVICE ROBERT ALIE Régie régionale de la santé et des services sociaux de la Montérégie 12, rue Beauregard,

Plus en détail

Essais cliniques de phase 0 : état de la littérature 2006-2009

Essais cliniques de phase 0 : état de la littérature 2006-2009 17 èmes Journées des Statisticiens des Centres de Lutte contre le Cancer 4 ème Conférence Francophone d Epidémiologie Clinique Essais cliniques de phase 0 : état de la littérature 2006-2009 Q Picat, N

Plus en détail

CE QU IL FAUT SAVOIR PARTICIPATION À UN ESSAI CLINIQUE SUR UN MÉDICAMENT

CE QU IL FAUT SAVOIR PARTICIPATION À UN ESSAI CLINIQUE SUR UN MÉDICAMENT CE QU IL FAUT SAVOIR PARTICIPATION À UN ESSAI CLINIQUE SUR UN MÉDICAMENT Sommaire Comment se fait la recherche sur un nouveau médicament? (page 1) A quoi sert la recherche sur un nouveau médicament? (page

Plus en détail

ALK et cancers broncho-pulmonaires. Laurence Bigay-Gamé Unité d oncologie thoracique Hôpital Larrey, Toulouse

ALK et cancers broncho-pulmonaires. Laurence Bigay-Gamé Unité d oncologie thoracique Hôpital Larrey, Toulouse ALK et cancers broncho-pulmonaires Laurence Bigay-Gamé Unité d oncologie thoracique Hôpital Larrey, Toulouse Toulouse, le 19 Février 2013 Adénocarcinomes : Lung Cancer Mutation Consortium Identification

Plus en détail

Artéfact en queue de comète à l échographie hépatique: un signe de maladie des voies biliaires intra-hépatiques

Artéfact en queue de comète à l échographie hépatique: un signe de maladie des voies biliaires intra-hépatiques Artéfact en queue de comète à l échographie hépatique: un signe de maladie des voies biliaires intra-hépatiques T Smayra, L Menassa-Moussa, S Slaba, M Ghossain, N Aoun Hôtel-Dieu de France, Université

Plus en détail

Qu avez-vous appris pendant cet exposé?

Qu avez-vous appris pendant cet exposé? Qu avez-vous appris pendant cet exposé? Karl VINCENT, DC IFEC 2010 Diplôme Universitaire de Méthodes en Recherche Clinique Institut de Santé Publique d Epidémiologie et de développement ISPED Bordeaux

Plus en détail

Programme-cadre et détail du programme des examens relatifs aux modules des cours de technologie, théorie professionnelle

Programme-cadre et détail du programme des examens relatifs aux modules des cours de technologie, théorie professionnelle Profil des compétences professionnelles Programme-cadre et détail du programme des examens relatifs aux modules des cours de technologie, théorie professionnelle Organisation pratique, lieux d organisation,

Plus en détail

COMMISSION NATIONALE D EVALUATION DES DISPOSITIFS MEDICAUX ET DES TECHNOLOGIES DE SANTE. AVIS DE LA COMMISSION 08 février 2011 CONCLUSIONS

COMMISSION NATIONALE D EVALUATION DES DISPOSITIFS MEDICAUX ET DES TECHNOLOGIES DE SANTE. AVIS DE LA COMMISSION 08 février 2011 CONCLUSIONS COMMISSION NATIONALE D EVALUATION DES DISPOSITIFS MEDICAUX ET DES TECHNOLOGIES DE SANTE AVIS DE LA COMMISSION 08 février 2011 Nom : Modèles et références retenus : Fabricant : Demandeur : Données disponibles

Plus en détail

Faut-il faire une corticothérapie prophylactique avant une césarienne élective pratiquée à 38 SA?

Faut-il faire une corticothérapie prophylactique avant une césarienne élective pratiquée à 38 SA? Faut-il faire une corticothérapie prophylactique avant une césarienne élective pratiquée à 38 SA? (césarienne élective = césarienne pratiquée "à froid", en dehors du travail) C Dageville réanimation néonatale,

Plus en détail

Hospital Anxiety and Depression Scale (HADS)

Hospital Anxiety and Depression Scale (HADS) dmt Risques psychosociaux : out ils d é va lua t ion FRPS 13 CATÉGORIE ATTEINTE À LA SANTÉ PHYSIQUE ET MENTALE Hospital Anxiety and Depression Scale (HADS) LANGEVIN V.*, FRANÇOIS M.**, BOINI S.***, RIOU

Plus en détail

Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Programme

Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Programme Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Programme Année universitaire / Academic Year: 2013 2014 A REMPLIR

Plus en détail

Optez pour une meilleure préparation canalaire avec la technologie Adaptive.

Optez pour une meilleure préparation canalaire avec la technologie Adaptive. Optez pour une meilleure préparation canalaire avec la technologie Adaptive. La préparation d un canal est souvent une opération complexe pour le dentiste. Le mouvement technologique Adaptive relève cependant

Plus en détail

L arthrose, ses maux si on en parlait!

L arthrose, ses maux si on en parlait! REF 27912016 INFORMER UPSA BROCH ARTHROSE V2 08-12.indd 2 30/08/12 11:48 Qu est-ce que l arthrose? L arthrose est une maladie courante des articulations dont la fréquence augmente avec l âge. C est une

Plus en détail

Instructions Mozilla Thunderbird Page 1

Instructions Mozilla Thunderbird Page 1 Instructions Mozilla Thunderbird Page 1 Instructions Mozilla Thunderbird Ce manuel est écrit pour les utilisateurs qui font déjà configurer un compte de courrier électronique dans Mozilla Thunderbird et

Plus en détail

ETABLISSEMENT D ENSEIGNEMENT OU ORGANISME DE FORMATION / UNIVERSITY OR COLLEGE:

ETABLISSEMENT D ENSEIGNEMENT OU ORGANISME DE FORMATION / UNIVERSITY OR COLLEGE: 8. Tripartite internship agreement La présente convention a pour objet de définir les conditions dans lesquelles le stagiaire ci-après nommé sera accueilli dans l entreprise. This contract defines the

Plus en détail

LOOPED THREAD OSTEOSUTURE: SMARTLOOP

LOOPED THREAD OSTEOSUTURE: SMARTLOOP Améliorer la consolidation des tubérosités grâce à une ostéosuture au fil boucle : le smartloop Improve tuberosity consolidation using a looped thread osteosuture: Smartloop T. Lascar, L. Obert Introduction

Plus en détail

Support Orders and Support Provisions (Banks and Authorized Foreign Banks) Regulations

Support Orders and Support Provisions (Banks and Authorized Foreign Banks) Regulations CANADA CONSOLIDATION CODIFICATION Support Orders and Support Provisions (Banks and Authorized Foreign Banks) Regulations Règlement sur les ordonnances alimentaires et les dispositions alimentaires (banques

Plus en détail

Pierre OLIVIER - Médecine Nucléaire

Pierre OLIVIER - Médecine Nucléaire Diplôme Universitaire Corrélations anatomo-physio-pathologiques en imagerie thoracique 25 mai 2011 Imagerie TEP et pathologie tumorale bronchique Pierre OLIVIER - Médecine Nucléaire Détection en coincidence

Plus en détail

Credit Note and Debit Note Information (GST/ HST) Regulations

Credit Note and Debit Note Information (GST/ HST) Regulations CANADA CONSOLIDATION CODIFICATION Credit Note and Debit Note Information (GST/ HST) Regulations Règlement sur les renseignements à inclure dans les notes de crédit et les notes de débit (TPS/ TVH) SOR/91-44

Plus en détail

Le traitement conservateur des tumeurs malignes des membres a largement remplacé les amputations

Le traitement conservateur des tumeurs malignes des membres a largement remplacé les amputations FACE A FACE Péroné vascularisé / Péroné non vascularisé Reconstruction après résection tumorale Dr NOURI / Dr BEN MAITIGUE SOTCOT Juin 2012 Le traitement conservateur des tumeurs malignes des membres a

Plus en détail

Le dropéridol n est pas un traitement à considérer pour le traitement de la migraine à l urgence

Le dropéridol n est pas un traitement à considérer pour le traitement de la migraine à l urgence Le dropéridol n est pas un traitement à considérer pour le traitement de la migraine à l urgence On pose fréquemment le diagnostic de migraine sévère à l urgence. Bien que ce soit un diagnostic commun,

Plus en détail

*smith&nephew IV3000 Pansement pour cathéters réactif à l'humidité. Le pansement idéal pour cathéters

*smith&nephew IV3000 Pansement pour cathéters réactif à l'humidité. Le pansement idéal pour cathéters Le pansement idéal pour cathéters IV3000 diminue sensiblement le risque d infection en maintenant sec le site de ponction. IV3000 est jusqu à 8 fois plus efficace en terme d évaporation d humidité que

Plus en détail

AIDE FINANCIÈRE POUR ATHLÈTES FINANCIAL ASSISTANCE FOR ATHLETES

AIDE FINANCIÈRE POUR ATHLÈTES FINANCIAL ASSISTANCE FOR ATHLETES AIDE FINANCIÈRE POUR ATHLÈTES FINANCIAL ASSISTANCE FOR ATHLETES FORMULAIRE DE DEMANDE D AIDE / APPLICATION FORM Espace réservé pour l utilisation de la fondation This space reserved for foundation use

Plus en détail

L axe 5 du Cancéropole Nord Ouest

L axe 5 du Cancéropole Nord Ouest L axe 5 du Cancéropole Nord Ouest Cancers, Individu id & Société L état des lieux d un pari Le Rapport Cordier 1 Biomarqueurs prédictifs 2 L axe 5 du Cancéropole Nord Ouest Cancers, Individu & Société

Plus en détail

ARTHROPLASTIES BILATÉRALES en UN ou DEUX TEMPS comparaison des risques périopératoires

ARTHROPLASTIES BILATÉRALES en UN ou DEUX TEMPS comparaison des risques périopératoires Dr Alain CAZENAVE chirurgien orthopédiste traumatologie sportive 52 rue du Dr calot 62600 BERCK/mer FRANCE tel: 03 21 89 20 44 site web : www.orthopale.org ARTHROPLASTIES BILATÉRALES en UN ou DEUX TEMPS

Plus en détail

Stéphane Lefebvre. CAE s Chief Financial Officer. CAE announces Government of Canada participation in Project Innovate.

Stéphane Lefebvre. CAE s Chief Financial Officer. CAE announces Government of Canada participation in Project Innovate. Stéphane Lefebvre CAE s Chief Financial Officer CAE announces Government of Canada participation in Project Innovate Montreal, Canada, February 27, 2014 Monsieur le ministre Lebel, Mesdames et messieurs,

Plus en détail

Annales du Contrôle National de Qualité des Analyses de Biologie Médicale

Annales du Contrôle National de Qualité des Analyses de Biologie Médicale Annales du Contrôle National de Qualité des Analyses de Biologie Médicale Plombémie Plombémie 07PLO1 ; 07PLO2 ; 07PLO3 et 07PLO4 2007 Edition : décembre 2008 Afssaps -143/147, Bd Anatole France F-93285

Plus en détail

IDENTITÉ DE L ÉTUDIANT / APPLICANT INFORMATION

IDENTITÉ DE L ÉTUDIANT / APPLICANT INFORMATION vice Direction des Partenariats Internationaux Pôle Mobilités Prrogrramme de bourrses Intterrnattiionalles en Mastterr (MIEM) Intterrnattiionall Mastterr Schollarrshiip Prrogrramme Sorrbonne Parriis Ciitté

Plus en détail

L évaluation de la qualité d un dispositif d apprentissage en ligne. Quelles traces mobiliser? Comment les interpréter?

L évaluation de la qualité d un dispositif d apprentissage en ligne. Quelles traces mobiliser? Comment les interpréter? L évaluation de la qualité d un dispositif d apprentissage en ligne. Quelles traces mobiliser? Comment les interpréter? François GEORGES (LabSET ULg) Séminaire Be-ODL 18 novembre 2013 1 Plan La qualité

Plus en détail

Photoactivatable Probes for Protein Labeling

Photoactivatable Probes for Protein Labeling Photoactivatable Probes for Protein Labeling THÈSE N O 4660 (2010) PRÉSENTÉE LE 26 MARS 2010 À LA FACULTÉ SCIENCES DE BASE LABORATOIRE D'INGÉNIERIE DES PROTÉINES PROGRAMME DOCTORAL EN CHIMIE ET GÉNIE CHIMIQUE

Plus en détail

Androids as an Experimental Apparatus: Why Is There an Uncanny Valley and Can we Exploit It?

Androids as an Experimental Apparatus: Why Is There an Uncanny Valley and Can we Exploit It? Androids as an Experimental Apparatus: Why Is There an Uncanny Valley and Can we Exploit It? Karl F. MacDorman 1 The Uncanny Valley : Le terme d Uncanny Valley a été défini par Masahiro Mori, chercheur

Plus en détail

SARM: Simulation of Absorption Refrigeration Machine

SARM: Simulation of Absorption Refrigeration Machine Revue des Energies Renouvelables Vol. 11 N 4 (2008) 587 594 SARM: Simulation of Absorption Refrigeration Machine S. Kherris 1*, M. Makhlouf 1 et A. Asnoun 2 1 Laboratoire des Matériaux et des Systèmes

Plus en détail

Cancer bronchique primitif: données épidémiologiques récentes

Cancer bronchique primitif: données épidémiologiques récentes Cancer bronchique primitif: données épidémiologiques récentes Pr Jean Trédaniel Service de pneumologie et oncologie thoracique Groupe Hospitalier Paris Saint Joseph Université Paris Descartes Sources Données

Plus en détail

TEST DE DÉTECTION DE LA PRODUCTION D INTERFÉRON γ POUR LE DIAGNOSTIC DES INFECTIONS TUBERCULEUSES

TEST DE DÉTECTION DE LA PRODUCTION D INTERFÉRON γ POUR LE DIAGNOSTIC DES INFECTIONS TUBERCULEUSES TEST DE DÉTECTION DE LA PRODUCTION D INTERFÉRON γ POUR LE DIAGNOSTIC DES INFECTIONS TUBERCULEUSES Classement NABM : non inscrit code : non codé DÉCEMBRE 2006 Service évaluation des actes professionnels

Plus en détail

Forthcoming Database

Forthcoming Database DISS.ETH NO. 15802 Forthcoming Database A Framework Approach for Data Visualization Applications A dissertation submitted to the SWISS FEDERAL INSTITUTE OF TECHNOLOGY ZURICH for the degree of Doctor of

Plus en détail

27 ème JOURNEE SPORT ET MEDECINE Dr Roukos ABI KHALIL Digne Les Bains 23 novembre 2013

27 ème JOURNEE SPORT ET MEDECINE Dr Roukos ABI KHALIL Digne Les Bains 23 novembre 2013 27 ème JOURNEE SPORT ET MEDECINE Dr Roukos ABI KHALIL Digne Les Bains 23 novembre 2013 Les Nouveaux Anticoagulants Oraux (NACO) dans la maladie thrombo embolique veineuse INTRODUCTION Thrombose veineuse

Plus en détail

Vers une approche Adaptative pour la Découverte et la Composition Dynamique des Services

Vers une approche Adaptative pour la Découverte et la Composition Dynamique des Services 69 Vers une approche Adaptative pour la Découverte et la Composition Dynamique des Services M. Bakhouya, J. Gaber et A. Koukam Laboratoire Systèmes et Transports SeT Université de Technologie de Belfort-Montbéliard

Plus en détail

Prise en Charge de la Fracture du Col du Fémur. Dr Christine LAFONT Service de Gériatrie C.H.U. TOULOUSE

Prise en Charge de la Fracture du Col du Fémur. Dr Christine LAFONT Service de Gériatrie C.H.U. TOULOUSE Prise en Charge de la Fracture du Col du Fémur Dr Christine LAFONT Service de Gériatrie C.H.U. TOULOUSE Épidémiologie «Problème médical, social et économique» Incidence : France : 50/100 000 hab. Augmente

Plus en détail

Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Program

Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Program Formulaire de candidature pour les bourses de mobilité internationale niveau Master/ Application Form for International Master Scholarship Program Année universitaire / Academic Year: 2013 2014 INTITULE

Plus en détail

Un nouveau standard d étanchéité pour une durée de vie accrue

Un nouveau standard d étanchéité pour une durée de vie accrue Un nouveau standard d étanchéité pour une durée de vie accrue Roulements rigides à billes munis de joints RSL et RSH Propriétés d étanchéité améliorées Faible frottement Durée de vie accrue Roulements

Plus en détail

Généralités. Aperçu. Introduction. Précision. Instruction de montage. Lubrification. Conception. Produits. Guides à brides FNS. Guides standards GNS

Généralités. Aperçu. Introduction. Précision. Instruction de montage. Lubrification. Conception. Produits. Guides à brides FNS. Guides standards GNS Généralités Aperçu Introduction Précision Instruction de montage Lubrification Conception page............................. 4............................. 5............................. 6.............................

Plus en détail

Essai Inter-groupe : FFCD UNICANCER FRENCH - GERCOR

Essai Inter-groupe : FFCD UNICANCER FRENCH - GERCOR CLIMAT - PRODIGE 30 Etude de phase III randomisée évaluant l Intérêt de la colectomie première chez les patients porteurs d un cancer colique asymptomatique avec métastases hépatiques synchrones non résécables

Plus en détail

biocer - système d implant dentaire

biocer - système d implant dentaire i n f o r m a t i o n s p r o d u i t s biocer - - implants dentaires combinés et concepts de butées en matériaux biologiquement fiables. Le dispositif exceptionnel des surfaces cerid - sont la conception

Plus en détail

Sujets présentés par le Professeur Olivier CUSSENOT

Sujets présentés par le Professeur Olivier CUSSENOT ANAMACaP Association Nationale des Malades du Cancer de la Prostate 17, bis Avenue Poincaré. 57400 SARREBOURG Sujets présentés par le Professeur Olivier CUSSENOT Place des nouvelles techniques d imagerie

Plus en détail

La chirurgie dans la PC

La chirurgie dans la PC La chirurgie dans la PC GF PENNECOT Marchant ou déambulant Grand handicapé Kinésith sithérapie Appareillage Médicaments Chirurgie neuro ortho LESION NEUROLGIQUE STABLE Spasticité Perte de la sélectivit

Plus en détail

IMMED Monitoring vidéo porté

IMMED Monitoring vidéo porté IMMED Monitoring vidéo porté L indexation vidéo au service du soin des personnes Projet financé par PEPS S2TI CNRS et des bourses BQR de l Université Bordeaux 1 1 Contexte Maladies et dépendances liées

Plus en détail

Graisses MULTIS COMPLEX. La haute technologie au service d environnements industriels extrêmes

Graisses MULTIS COMPLEX. La haute technologie au service d environnements industriels extrêmes Graisses La haute technologie au service d environnements industriels extrêmes , le haut de gamme des graisses multi-fonctions n SHD Graisses synthétiques PAO Températures très basses à fortes Vitesses

Plus en détail

Information pour le patient

Information pour le patient Information pour le patient Plus de qualité de vie avec des implants dentaires SWISS PRECISION AND INNOVATION. 1 Table des matières Quelle est l importance des dents pour notre qualité de vie? 4 Qu est-ce

Plus en détail

Avis 29 mai 2013. XYZALL 5 mg, comprimé B/14 (CIP : 34009 358 502 4-9) B/28 (CIP : 34009 358 505 3-9) Laboratoire UCB PHARMA SA.

Avis 29 mai 2013. XYZALL 5 mg, comprimé B/14 (CIP : 34009 358 502 4-9) B/28 (CIP : 34009 358 505 3-9) Laboratoire UCB PHARMA SA. COMMISSION DE LA TRANSPARENCE Avis 29 mai 2013 XYZALL 5 mg, comprimé B/14 (CIP : 34009 358 502 4-9) B/28 (CIP : 34009 358 505 3-9) Laboratoire UCB PHARMA SA DCI Code ATC (2012) Motif de l examen Liste

Plus en détail